Applications
- Drinking water
- Municipal wastewater
- Industrial wastewater
- hydrocarbons monitorning
- Oil refinement/transportation
- Pollution monitoring of surface waters
Description
Online probe for the determination of organic load, COD, BOD, SAC254. The sensor is based on the UV photometric absorbance measurement at 254 nm of active species dissolved or suspended in water. The external probe is installed in a sample reservoir and measurements are taken on the circulating sample flow, guaranteeing a short response time. The measurement of the sensor can be correlated with various analytical parameters such as COD, TOC, BOD, SAC254. Applications includes industrial and municipal wastewater processing, environmental monitoring, potable water.
Method
The principle of measurement is based on the determination of the absorbance of organic species dissolved in water. The low range version of the probe is indicated for organic loads with a COD(KHPeq) between 0 and 370 mg/l.
Features
Measured parameter | COD |
Wavelength | 254 nm |
Demineralized water | Not required |
Reagents and standards
- No reagents required.
- See standard solutions for UVO LOW analyzer.
The principle of measurement is based on the determination of the absorbance of organic species dissolved in water. The high range version of the probe is indicated for organic loads with a COD(KHPeq) between 0 and 1000 mg/l.
Features
Measured parameter | COD |
Wavelength | 254 nm |
Demineralized water | Not required |
Reagents and standards
- No reagents required.
- See standard solutions for UVO HIGH analyzer.
Technical Specifications
Model | 3S-UVO |
Sampling Mode | Continuous. |
Sample | Pressure: Atmospheric Temperature: 5 ‐ 50°C (41 ‐ 122°F) Flow Rate: 80 to 500 mL/min Connection: 6 mm (¼-in.) |
Drain |
Pressure-free drain Connection: 12 mm (½-in.) |
Input Streams | 2 |
Material | Stainless steel 316L. |
Dimensions (H x W x D) | 176 x 50 x 50 mm (6.9 x 2.0 x 2.0 in) |
Weight | 1 kg (2 lbs) |
Power Supply |
Input Voltage: 115 VAC,230 VAC Power consumption: max. 80 W |
Output signals |
2 x 4-20 mA analog outputs Modbus via RS485,Ethernet |
Alarms | 2 SPDT programmable voltage-free relays |
Digital Input | None |
Ambient Temperature | 5 - 50 °C (41 - 122 °F) |
Ambient Humidity | 100 % RH |
Protection Grade | IP68 |